Evaluating the Agreement between Oral, Armpit, and Ear Temperature Readings during Physical Activities in an Outdoor Setting
Yuanzhe Zhao,
Leonardo de Almeida e Bueno,
David A. Holdsworth and
Jeroen H. M. Bergmann ()
Additional contact information
Yuanzhe Zhao: Department of Engineering Science, University of Oxford, Parks Road, Oxford OX1 3PJ, UK
Leonardo de Almeida e Bueno: Department of Engineering Science, University of Oxford, Parks Road, Oxford OX1 3PJ, UK
David A. Holdsworth: Oxford University Hospitals NHS Foundation Trust, Oxford OX3 9DU, UK
Jeroen H. M. Bergmann: Department of Engineering Science, University of Oxford, Parks Road, Oxford OX1 3PJ, UK
IJERPH, 2024, vol. 21, issue 5, 1-11
Abstract:
Accurate body temperature measurement is essential for monitoring and managing safety during outdoor activities. Physical activities are an essential consideration for public health, with sports taking up an important proportion of these. Athletes’ performances can be directly affected by body temperature fluctuations, with overheating or hypothermia posing serious health risks. Monitoring these temperatures allows coaches and medical staff to make decisions that enhance performance and safety. Traditional methods, like oral, axillary, and tympanic readings, are widely used, but face challenges during intense physical activities in real-world environments. This study evaluated the agreement, correlation, and interchangeability of oral, axillary, and tympanic temperature measurements in outdoor exercise conditions. Systems developed for specific placements might generate different sensor readouts. Conducted as an observational field study, it involved 21 adult participants (11 males and 10 females, average age 25.14 ± 5.80 years) that underwent the Yo-Yo intermittent recovery test protocol on an outdoor court. The main outcomes measured were the agreement and correlation between temperature readings from the three methods, both before and after exercise. The results indicate poor agreement between the measurement sites, with significant deviations observed post-exercise. Although the Spearman correlation coefficients showed consistent temperature changes post-exercise across all methods, the standard deviations in the pairwise comparisons exceeded 0.67 °C. This study concluded that widely used temperature measurement methods are challenging to use during outdoor exercises and should not be considered interchangeable. This variability, especially after exercise, underscores the need for further research using gold standard temperature measurement methods to determine the most suitable site for accurate readings. Care should thus be taken when temperature screening is done at scale using traditional methods, as each measurement site should be considered within its own right.
Keywords: body temperature; oral temperature; tympanic temperature; axillary temperature (search for similar items in EconPapers)
JEL-codes: I I1 I3 Q Q5 (search for similar items in EconPapers)
Date: 2024
References: View references in EconPapers View complete reference list from CitEc
Citations:
Downloads: (external link)
https://www.mdpi.com/1660-4601/21/5/595/pdf (application/pdf)
https://www.mdpi.com/1660-4601/21/5/595/ (text/html)
Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.
Export reference: BibTeX
RIS (EndNote, ProCite, RefMan)
HTML/Text
Persistent link: https://EconPapers.repec.org/RePEc:gam:jijerp:v:21:y:2024:i:5:p:595-:d:1388516
Access Statistics for this article
IJERPH is currently edited by Ms. Jenna Liu
More articles in IJERPH from MDPI
Bibliographic data for series maintained by MDPI Indexing Manager ().