Management of Deep Brain Stimulator Battery Failure: Battery Estimators, Charge Density, and Importance of Clinical Symptoms

Kaihan Fakhar, Erin Hastings, Christopher R Butson, Kelly D Foote, Pam Zeilman and Michael S Okun

PLOS ONE, 2013, vol. 8, issue 3, 1-8

Abstract: Objective: We aimed in this investigation to study deep brain stimulation (DBS) battery drain with special attention directed toward patient symptoms prior to and following battery replacement. Background: Previously our group developed web-based calculators and smart phone applications to estimate DBS battery life (http://mdc.mbi.ufl.edu/surgery/dbs-battery-estimator). Methods: A cohort of 320 patients undergoing DBS battery replacement from 2002–2012 were included in an IRB approved study. Statistical analysis was performed using SPSS 20.0 (IBM, Armonk, NY). Results: The mean charge density for treatment of Parkinson’s disease was 7.2 µC/cm2/phase (SD = 3.82), for dystonia was 17.5 µC/cm2/phase (SD = 8.53), for essential tremor was 8.3 µC/cm2/phase (SD = 4.85), and for OCD was 18.0 µC/cm2/phase (SD = 4.35). There was a significant relationship between charge density and battery life (r = −.59, p

Date: 2013
References: View complete reference list from CitEc
Citations:

Downloads: (external link)
https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0058665 (text/html)
https://journals.plos.org/plosone/article/file?id= ... 58665&type=printable (application/pdf)

Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.

Export reference: BibTeX RIS (EndNote, ProCite, RefMan) HTML/Text

Persistent link: https://EconPapers.repec.org/RePEc:plo:pone00:0058665

DOI: 10.1371/journal.pone.0058665

Access Statistics for this article

More articles in PLOS ONE from Public Library of Science
Bibliographic data for series maintained by plosone ().

 
Page updated 2025-03-22
Handle: RePEc:plo:pone00:0058665