Accuracy and precision of ScvO2 measured with the CeVOX-device: A prospective study in patients with a wide variation of ScvO2-values
Alexander Herner,
Bernhard Haller,
Ulrich Mayr,
Sebastian Rasch,
Lea Offman,
Roland Schmid and
Wolfgang Huber
PLOS ONE, 2018, vol. 13, issue 4, 1-16
Abstract:
Introduction: Central-venous oxygen saturation (ScvO2) is a key parameter of hemodynamic monitoring and has been suggested as therapeutic goal for resuscitation. Several devices offer continuous monitoring features. The CeVOX-device (Pulsion Medical Systems) uses a fibre-optic probe inserted through a conventional central-venous catheter (CVC) to obtain continuous ScvO2. Objectives: Since there is a lack of studies validating the CeVOX, we prospectively analyzed data from 24 patients with CeVOX-monitoring. To increase the yield of lower ScvO2-values, 12 patients were equipped with a femoral CVC. Methods: During the 8h study period ScvO2_CeVOX was documented immediately before withdrawal of blood to measure ScvO2 by blood gas analysis (ScvO2_BGA) 6min, 1h, 4h, 5h and 8h after the initial calibration. No further calibrations were performed. Results: In patients with jugular CVC (primary endpoint; 60 measurements), bias, lower and upper limits of agreement (LLOA; ULOA) and percentage error (PE) of the estimate of ScvO2 (ScvO2_CeVOX_jug) were acceptable with 0.45%, -13.0%, 13.9% and 16.6%, respectively. Conclusion: These findings suggest that the CeVOX device is capable to detect stability or instability of ScvO2_BGA. ScvO2_CeVOX accurately estimates ScvO2_BGA in case of stable values. However, intermittent measurement of ScvO2_BGA and re-calibration should be performed in case of substantial changes in ScvO2_CeVOX compared to baseline. Therefore, continuous measurement of ScvO2 with the CeVOX cannot replace ScvO2_BGA in instable patients. On the other hand, CeVOX might be useful for the monitoring of stable patients as a pre-test tool for more differentiated monitoring in case of changes in ScvO2_CeVOX.
Date: 2018
References: View complete reference list from CitEc
Citations:
Downloads: (external link)
https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0192073 (text/html)
https://journals.plos.org/plosone/article/file?id= ... 92073&type=printable (application/pdf)
Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.
Export reference: BibTeX
RIS (EndNote, ProCite, RefMan)
HTML/Text
Persistent link: https://EconPapers.repec.org/RePEc:plo:pone00:0192073
DOI: 10.1371/journal.pone.0192073
Access Statistics for this article
More articles in PLOS ONE from Public Library of Science
Bibliographic data for series maintained by plosone ().