Towards unambiguous reporting of complications related to deep brain stimulation surgery: A retrospective single-center analysis and systematic review of the literature
Katja Engel,
Torge Huckhagel,
Alessandro Gulberti,
Monika Pötter-Nerger,
Eik Vettorazzi,
Ute Hidding,
Chi-un Choe,
Simone Zittel,
Hanna Braaß,
Peter Ludewig,
Miriam Schaper,
Kara Krajewski,
Christian Oehlwein,
Katrin Mittmann,
Andreas K Engel,
Christian Gerloff,
Manfred Westphal,
Christian K E Moll,
Carsten Buhmann,
Johannes A Köppen and
Wolfgang Hamel
PLOS ONE, 2018, vol. 13, issue 8, 1-29
Abstract:
Background and objective: To determine rates of adverse events (AEs) related to deep brain stimulation (DBS) surgery or implanted devices from a large series from a single institution. Sound comparisons with the literature require the definition of unambiguous categories, since there is no consensus on the reporting of such AEs. Patients and methods: 123 consecutive patients (median age 63 yrs; female 45.5%) treated with DBS in the subthalamic nucleus (78 patients), ventrolateral thalamus (24), internal pallidum (20), and centre médian-parafascicular nucleus (1) were analyzed retrospectively. Both mean and median follow-up time was 4.7 years (578 patient-years). AEs were assessed according to three unambiguous categories: (i) hemorrhages including other intracranial complications because these might lead to neurological deficits or death, (ii) infections and similar AEs necessitating the explantation of hardware components as this results in the interruption of DBS therapy, and (iii) lead revisions for various reasons since this involves an additional intracranial procedure. For a systematic review of the literature AE rates were calculated based on primary data presented in 103 publications. Heterogeneity between studies was assessed with the I2 statistic and analyzed further by a random effects meta-regression. Publication bias was analyzed with funnel plots. Results: Surgery- or hardware-related AEs (23) affected 18 of 123 patients (14.6%) and resolved without permanent sequelae in all instances. In 2 patients (1.6%), small hemorrhages in the striatum were associated with transient neurological deficits. In 4 patients (3.3%; 0.7% per patient-year) impulse generators were removed due to infection. In 2 patients electrodes were revised (1.6%; 0.3% per patient-year). There was no lead migration or surgical revision because of lead misplacement. Age was not statistically significant different (p>0.05) between patients affected by AEs or not. AE rates did not decline over time and similar incidences were found among all patients (423) implanted with DBS systems at our institution until December 2016. A systematic literature review revealed that exact AE rates could not be determined from many studies, which could not be attributed to study designs. Average rates for intracranial complications were 3.8% among studies (per-study analysis) and 3.4% for pooled analysis of patients from different studies (per-patient analysis). Annual hardware removal rates were 3.6 and 2.4% for per-study and per-patient analysis, respectively, and lead revision rates were 4.1 and 2.6%, respectively. There was significant heterogeneity between studies (I2 ranged between 77% and 91% for the three categories; p
Date: 2018
References: View references in EconPapers View complete reference list from CitEc
Citations:
Downloads: (external link)
https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0198529 (text/html)
https://journals.plos.org/plosone/article/file?id= ... 98529&type=printable (application/pdf)
Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.
Export reference: BibTeX
RIS (EndNote, ProCite, RefMan)
HTML/Text
Persistent link: https://EconPapers.repec.org/RePEc:plo:pone00:0198529