Transportable system for on‐site calibration of permafrost temperature sensors
Andrea Merlone,
Francesca Sanna,
Graziano Coppa,
Laura Massano and
Chiara Musacchio
Permafrost and Periglacial Processes, 2020, vol. 31, issue 4, 610-620
Abstract:
Evaluating the degradation of permafrost is a major challenge in understanding global warming and its impact on the cryosphere. The Global Cryosphere Watch is promoting actions towards data quality and traceability, to achieve comparability of observations from different permafrost stations. In response to this, a transportable system for on‐site calibrations of permafrost temperature sensors was studied, developed and tested in the field, within the project MeteoMet. The system, here described, allows users to establish metrological traceability to permafrost temperature profiles, by performing the calibration on‐site, even in remote or high‐elevation areas, in realistic conditions. A field campaign at 3,000 m elevation to test the system's performance and practical use is also reported. Overall calibration uncertainty in the field accounted for
Date: 2020
References: View complete reference list from CitEc
Citations:
Downloads: (external link)
https://doi.org/10.1002/ppp.2063
Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.
Export reference: BibTeX
RIS (EndNote, ProCite, RefMan)
HTML/Text
Persistent link: https://EconPapers.repec.org/RePEc:wly:perpro:v:31:y:2020:i:4:p:610-620
Access Statistics for this article
More articles in Permafrost and Periglacial Processes from John Wiley & Sons
Bibliographic data for series maintained by Wiley Content Delivery ().