Grid Monte Carlo Simulation of a Medical Linear Accelerator
Didi Samir,
Mustapha Zerfaoui,
Abdelilah Moussa,
Yassine Benkhouya and
Mehdi el Ouartiti
European Journal of Engineering and Technology Research, 2018, vol. 3, issue 12, 40-43
Abstract:
A full grid simulation of the head of an Elekta Synergy Platform medical linear accelerator is performed using the Geant4 Monte Carlo platform. The simulation includes all components of the accelerator head and a homogeneous water phantom. Results in terms of depth doses and lateral dose profiles are presented for 6 MV photon beam with the 10x10 cm2 reference field size at 100 cm distance from the source. Overall, a good agreement with the measured dose data is achieved with a precision better than 0.93% and 2.63% for the depth dose profile and lateral dose profiles respectively.
Keywords: Grid Computing; Linac; Radiotherapy; Geant4. (search for similar items in EconPapers)
Date: 2018
References: View complete reference list from CitEc
Citations:
Downloads: (external link)
https://eu-opensci.org/index.php/ejeng/article/view/61001 Abstract page (text/html)
https://eu-opensci.org/index.php/ejeng/article/download/61001/11990 Full text (application/pdf)
Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.
Export reference: BibTeX
RIS (EndNote, ProCite, RefMan)
HTML/Text
Persistent link: https://EconPapers.repec.org/RePEc:epw:ejeng0:v:3:y:2018:i:12:id:61001
DOI: 10.24018/ejeng.2018.3.12.1001
Access Statistics for this article
More articles in European Journal of Engineering and Technology Research from European Open Science
Bibliographic data for series maintained by Support ().