Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions
Juan D N Diaz de Pool,
Sjoerd A A Van Den Berg,
Gonneke S K Pilgram,
Bartholomeus E P B Ballieux and
Lucia A J Van Der Westerlaken
PLOS ONE, 2018, vol. 13, issue 11, 1-10
Abstract:
Measurement of pH in IVF-media using the blood gas analyzer (BGA) requires validation, because IVF-media is outside the intended scope of the BGA. To determine whether the Siemens Rapidpoint 500 BGA is suitable for pH measurements in IVF-media this study will validate the BGA and assess its accuracy. In this method comparison study, the pH of over three hundred IVF-media samples was measured with the BGA and a pH electrode (Hanna pH checker). The precision of both the BGA and the pH electrode were excellent (coefficient variation
Date: 2018
References: View complete reference list from CitEc
Citations:
Downloads: (external link)
https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0206707 (text/html)
https://journals.plos.org/plosone/article/file?id= ... 06707&type=printable (application/pdf)
Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.
Export reference: BibTeX
RIS (EndNote, ProCite, RefMan)
HTML/Text
Persistent link: https://EconPapers.repec.org/RePEc:plo:pone00:0206707