Finite element analysis of Stryker Xia pedicle screw in artificial bone samples with and without supplemental cement augmentation
Ferris M. Pfeiffer,
Theodore J. Choma and
Rebecca Kueny
Computer Methods in Biomechanics and Biomedical Engineering, 2015, vol. 18, issue 13, 1459-1467
Abstract:
A validated, using in vitro biomechanical testing, finite element model was used to evaluate the affects of (1) cement augmentation and (2) an intact posterior cortex in osteoporotic bone. The presence of augmentation and/or a posterior cortical cortex increased the stabilization of the pedicle screw 2–5 fold. Placement of cement influenced failure load and toggle; with distal placement having the largest increase in failure load and decrease in cephalad–caudad toggle. The presence of posterior cortex caused a decrease in the amount of toggle, a proximal shift of the center of rotation and an increase in the maximum failure force.
Date: 2015
References: View complete reference list from CitEc
Citations:
Downloads: (external link)
http://hdl.handle.net/10.1080/10255842.2014.916699 (text/html)
Access to full text is restricted to subscribers.
Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.
Export reference: BibTeX
RIS (EndNote, ProCite, RefMan)
HTML/Text
Persistent link: https://EconPapers.repec.org/RePEc:taf:gcmbxx:v:18:y:2015:i:13:p:1459-1467
Ordering information: This journal article can be ordered from
http://www.tandfonline.com/pricing/journal/gcmb20
DOI: 10.1080/10255842.2014.916699
Access Statistics for this article
Computer Methods in Biomechanics and Biomedical Engineering is currently edited by Director of Biomaterials John Middleton
More articles in Computer Methods in Biomechanics and Biomedical Engineering from Taylor & Francis Journals
Bibliographic data for series maintained by Chris Longhurst ().