EconPapers    
Economics at your fingertips  
 

Development of reduced volume endosseous cuspid tooth implant using topology optimization by SIMP technique for improved osseointegration

Priyanshu Soni, Parnika Shrivastava and Sanjay Kumar Rai

Computer Methods in Biomechanics and Biomedical Engineering, 2024, vol. 27, issue 16, 2362-2376

Abstract: The article aims to design and develop a topology-optimized endosseous cuspid tooth implant of the maxilla region. The manuscript presents a numerical analysis of the resulting von Mises stresses and effective strain resulting in the topology-optimized implant with occlusal loading of 110 N. Solid Isotropic Material with Penalization (SIMP) method is employed for topology optimization and four different models, namely model-1, model-2, model-3, and model-4, are developed based on volume reduction rates of 8%, 16%, 24%, and 32%, respectively. FEA results highlight that the maximum stress and strain in the screw increases with volume reduction rates. The comparative analyses of the resulting stresses in the compact and cancellous bone along with the strain in the screw led to the conclusion that model-1, model-2, and model-3 resulted in moderate stresses on compact and cancellous bone compared to the original model of the implant. However, the screw and bones are subjected to maximum stress and strain in the model-4. The study concludes that model-2, with 16% reduced volume and 14.2% reduced mass as compared to the original implant, may be considered as the optimized design of the model. The resulting model offers a significant reduction in the weight and volume with a minor increase in effective stress and strain without negatively impacting the functionality and bio-mechanical performance of the implant. The optimized dental implant prototype is also fabricated as a proof of concept by the Fused Deposition Modelling process.

Date: 2024
References: Add references at CitEc
Citations:

Downloads: (external link)
http://hdl.handle.net/10.1080/10255842.2023.2279939 (text/html)
Access to full text is restricted to subscribers.

Related works:
This item may be available elsewhere in EconPapers: Search for items with the same title.

Export reference: BibTeX RIS (EndNote, ProCite, RefMan) HTML/Text

Persistent link: https://EconPapers.repec.org/RePEc:taf:gcmbxx:v:27:y:2024:i:16:p:2362-2376

Ordering information: This journal article can be ordered from
http://www.tandfonline.com/pricing/journal/gcmb20

DOI: 10.1080/10255842.2023.2279939

Access Statistics for this article

Computer Methods in Biomechanics and Biomedical Engineering is currently edited by Director of Biomaterials John Middleton

More articles in Computer Methods in Biomechanics and Biomedical Engineering from Taylor & Francis Journals
Bibliographic data for series maintained by Chris Longhurst ().

 
Page updated 2025-03-20
Handle: RePEc:taf:gcmbxx:v:27:y:2024:i:16:p:2362-2376